Home > Product:Medical > Medical > Medical > Adsorptive Apheresis Monitor Adamonitor SC

Adsorptive Apheresis Monitor Adamonitor SC

Adamonitor is the Adsorptive Apheresis Monitor used with Adacolumn (a granulocyte and monocyte/macrophage adsorption device, by JIMRO Co.,  Ltd.). Adamonitor is an adsorptive apheresis device exclusively for apheresis at a low flow rate.
Manufacturing Notification No. (Japan): 22800BZX00360000

Features

Features
  • Capable of maintaining a stable flow rate even at a low flow rate.
  • Safety mechanism activated when various sensors detect abnormalities.

Specifications

Specifications
  Specifications of Device
Rated voltage 100-240 V AC
Rated frequency 50/60 Hz
Power consumption 200 VA
Dimensions 230(W) × 328(D) × 515(H) mm
Weight Approx.11kg
 ★Blood Pump
Pump type Two-roller pump
Present flow rate range Circulation stage: 10 to 50 mL / min
Preparation stage: 10 to 150 mL / min
Time setup range 1 to 180 minutes
Safety mechanism Pump cover operation sensor
Rotation Sensor
 ★Syringe Pump
Present flow rate 1 to 20 mL / h
Safety mechanism Syringe sensor
Syringe movement sensor
Syringe completion sensor
Plunger set sensor
Plunger overload sensor
Feed screw and nut of opening and closing sensor
Applicable syringe TERUMO, NIPRO, JMS, TOP
(20mL luer-locking type plastic syringe)
 ★Sensor
●Normal Saline Solution Outage Sensor
Detection system Ultrasonic system
●Bubble Sensor
Detection system Ultrasonic system
Detection sensibility Closed cell of 0.1 mL or over
Safety mechanism When the bubble sensor detects bubbles, the blood
pump is immediately shut down, the bubble circuit is
closed by the clamper, the abnormality is reported
with warning buzzer, warning lamp, and monitor display.
Almost no continuous air infusion measured
(much lower than the value of 0.03mL/min).
●Internal Pressure of Venous Circuit
Measurement display range ‐13.3 to 53.3 kPa(-100 to 400 mmHg)
Venous pressure warning
setup range
‐13.3 to 53.3 kPa(-100 to 400 mmHg)
Safety mechanism When the pressure in the dip chamber or the venous
circuit exceeds the venous pressure warning setup
range, the blood pump is stopped automatically, and
the abnormality is reported with the warning buzzer,
warning lamp.
●Pillow Sensor (Negative Arterial Pressure Sensor)
Detection system Mechanical switch
Safety mechanism When the abnormality (the pressure in the pillow
becomes negative and the pillow shrinks due to
blockage in blood vessel, etc) is detected by the
negative arterial pressure sensor, the blood pump is
immediately stopped, the blood circuit is closed by
the clamper, and the abnormality is reported with the
warning buzzer, warning lamp.

Related Information

 Safety and Performance Information

  [ Instructions Download ]

Page Top

X LinkdIn Youtube

Connect with us on social media